Instrument name:In vitro tissue water bath system
Specification and model:159928
Manufacturer:Radnoti
Application:artificial blood vessel pressure testing and antihypertensive phamaceuticals research.
Main functions of the instrument:
The main test objects include isolated muscles, arterial vascular rings and vascular strips. After eliminating the regulatory effects of nerves and body fluids, different conditions can be controled by the system. Changes of pressure and tension in isolated tissues can be exactly detected.
Technical specifications of the instrument:
1. Equipped with eight tissue baths and tension transducers;
2. Equipped with high-power heating circulation pump;
3. The system adopts modular design;
4. The insulation system adopts a water jacket irrigation conduit and container;
5. The system glassware is made of borosilicate glass;
6. Number of channels for 8-channel physiological recorder: 8;
7. Type: 8 BNC interfaces, 4 spare differential Pod interfaces (8-pin DIN);
8. Input voltage: ± 15 V;
9. Filtering: 1Hz-1kHz (in 2:5:10 steps); 2 kHz,25kHz;
10. Coupling: DC or 0.15 Hz (software selection);
11. Supply voltage: ± 5 V;
12. Maximum current: 50 mA;
13. Connection method: 2-wire I2C;
14. Interface type: 8-pin DIN;
15. Sampling ADC accuracy: 16 bits;
16. Maximum sampling rate: single channel or dual channel 200K; 3-4 channels 100K; 5-8 channel 40K;
17. Environmental temperature: 16 to 26 ℃;
18. Relative humidity: 40-80%.
Person in charge: Luxi Dong、Qiuping Zhang
Telephone: 022-85358229
E-mail: donglx@nankai.edu.cn
zhangqp@nankai.edu.cn
Address:D121-1, ComprehensiveLaboratory Building, Jinnan New Campus, Nankai University